Product Details
Place of Origin: Shanghai, China
Brand Name: Bing Wire & Cable
Certification: ISO, RoHS
Model Number: Medical RF Ablation Cable
Payment & Shipping Terms
Minimum Order Quantity: 500m
Price: $0.01 / m
Delivery Time: 7-15 work days
Payment Terms: T/T
Supply Ability: 1000m 7 work days
|
Type:
|
Medical RF Ablation Cable
|
Type-T Thermocouple Conductor:
|
Pure Copper Wire & Constantan Wire
|
Insulation:
|
FEP
|
RF Coaxial Cable:
|
19-strand Tinned Copper
|
Outer Jacket:
|
Silicone
|
|
Type:
|
Medical RF Ablation Cable
|
|
Type-T Thermocouple Conductor:
|
Pure Copper Wire & Constantan Wire
|
|
Insulation:
|
FEP
|
|
RF Coaxial Cable:
|
19-strand Tinned Copper
|
|
Outer Jacket:
|
Silicone
|
Custom 3-Core Construction | Blue Silicone or DEHP-Free PVC | 2.6 mm OD | 1.5 kV
In advanced minimally invasive surgeries, such as Radiofrequency (RF) or Microwave Ablation, delivering precise energy while simultaneously monitoring tissue temperature is critical for patient safety and efficacy. The Custom Medical RF Ablation Hybrid Cable is engineered to solve this exact challenge. By integrating a high-current 0.8 mm RF coaxial line and a highly sensitive dual-wire T-type thermocouple circuit into a single 2.6 mm outer diameter, this composite cable streamlines device architecture. It features a biocompatible, ultra-flexible blue medical-grade Silicone (or DEHP-free PVC) jacket, designed to withstand harsh sterilization processes while providing the flexibility required for complex surgical navigation. It is fully compliant with the most stringent medical safety and EMC standards.
Integrated Dual-Function Design – Combines RF energy transmission and real-time temperature monitoring in a single micro-cable, optimizing internal device space.
Ultra-Flexible & Biocompatible Jacket – Available in DEHP-free medical PVC (55-60A) or medical-grade blue silicone. Resists repeated bending, chemical disinfection, and autoclave sterilization.
High-Voltage Safety – Withstands AC 1.5 kV for 5 minutes without breakdown, ensuring absolute electrical isolation during high-frequency surgical procedures.
Precision Thermal Sensing – Utilizes 0.2 mm T-type thermocouple wires (Copper/Constantan) insulated with FEP, meeting GB/T4990 TX precision-grade standards.
Low-Resistance RF Conductor – Features a 19-strand tinned copper inner conductor, balancing low DC resistance (< 0.125 Ω/m) with exceptional flex-fatigue resistance.
Strict Medical Compliance – Designed and tested to meet GB 9706.1-2020, GB 9706.202-2021 (High-Frequency Surgical Equipment), and YY 0505-2012 (EMC) requirements.
| Component | Material and Construction | Color / Dimension |
|---|---|---|
| Outer Jacket | Blue Medical Silicone or DEHP-Free PVC | Ø 2.6 ± 0.1 mm |
| RF Coaxial Cable | 19-strand tinned copper (0.10 mm) | Ø 0.8 mm |
| Thermocouple 1 | T-type Copper wire + FEP | Ø 0.4 mm / Red |
| Thermocouple 2 | T-type Constantan wire + FEP | Ø 0.4 mm / White |
| Parameter | Specification |
|---|---|
| Outer Diameter | 2.6 ± 0.1 mm |
| Insulation Withstand Voltage | AC 1.5 kV / 5 minutes |
| RF Conductor Resistance | < 0.125 Ω/m (at 20°C) |
| Thermocouple Type | T-Type (Copper/Constantan), Precision Grade |
| Jacket Hardness | 55-60A |
| Applicable Standards | GB 9706.1-2020, GB 9706.202-2021, YY 0505-2012 |
Designing cables for RF ablation is notoriously difficult: high-frequency energy easily interferes with the delicate millivolt signals of a thermocouple. Our engineering team specializes in micro-coaxial and multi-core medical composite cables. We carefully control the cabling lay, insulation thickness, and shielding options to minimize electromagnetic interference (EMI) and pass stringent YY 0505 EMC testing. We support full product development—from structural design based on your drawings, to prototype sampling, and mass production.
Radiofrequency Ablation (RFA) – Liver, thyroid, and tumor ablation catheters.
Microwave Ablation (MWA) – High-frequency energy delivery with thermal sensing.
Minimally Invasive Surgical Instruments – Space-constrained medical probes.
High-Frequency Signal Transmission Equipment – Minimally invasive surgical probes.
Advanced Electrosurgical Units (ESU) – Precision energy delivery and safety monitoring.